Calibrator 3200 is the novelty, which replaced model 2100. The distinctive special features of this instrument are the renovated [ergonomichnyy] housing and the extended functional possibilities.
Complex calibration - The testers of isolation to 5 kV - BONDS devices of shielding turning off - The ohmmeters - The voltmeters - Testers for checking the devices of the grounding - The automatic calibration according to ISO 9000 s ON ProCal
Additional functions The functional possibilities of the calibrator of a series of 2100 at will customers are enlarged due to the options, which make it possible to carry out the calibration of ohmmeters, the auto-test of chain as far as two additional outputs for calibrating the bridges of relation. The calibration of instruments for measuring the voltage of sparking and current of spark is ensured.
ProCal - software, which makes it possible to conduct the complex calibration of equipment completely before the automatic regime.
3200 can be used both the laboratory calibrator and as far as movable, for the work due to the field conditions. Interface RS 232 makes it possible to work with portable PK. Characteristics
Stress is the nourishment: 230 [v] (- 6%, +10%), 50 Hz Required power: 28 W Interface: RS-232 (USB option) EMC - correspondence: EN50081, EN1010 The operating temperature: 0… 50°C with the complete accuracy. Warmup time: to 20 min. Storage temperature: -5°C… of 60°C. Relative humidity of the environment: to 80% with 30°C. Overall dimensions: 45 × 44 × 14 cm Mass of the instrument: 10 kgf
Are given below the technical characteristics of calibrator, united about the functional designation.
CALIBRATION OF THE TESTERS OF THE BREAK OF THE ELECTRICAL CHAINS Range of the adjustable resistors: 0,1 - 20 Ohm, permission 0,01 Ohm. Fixed resistances: 100 Ohm, 1 kilohms Standard accuracy: 1%±0,025 Ohm Improved accuracy (option): 0,25%±0,025[Om] Range of the measurement of the current of leakages (load 1 Ohm): From 0 to 320 mA Accuracy across the channel of the current: 1,3%±6 of [emr].
CALIBRATION OF THE TESTERS OF THE ISOLATION Range of the resistances: From 0 to 2 G (to 10 G - on the option), the permission of 10[kOm]. Accuracy: - 0,3% (from 0 to 5 M); 0,1% - option; - 3% (from by 5[Mom] to 2 G); 1% - option.
- 5% from 2[GOm] to 10 G
Ranges of the test stresses: 50 beside, 100 before, 250 beside, 500 before, 1 kV Ranges of the measurement of the current: 0,5 mA; 1 mA Accuracy (on the current and the voltage): 1%±8 of [emr]. Ranges of the generation of the voltage of alternating current: 100, 200, 230, 300 and 400 v. Accuracy of the generation of the stress: 0,2±1 of [emr].
CALIBRATION OF TESTERS for checking THE BONDS (devices of shielding turning off) Response time: from 20 ms to 5 number systems by permission 10 ms. Accuracy of the switching: ±0,7 ms. Range of the generation of the current: from 3 mA to 3a. Accuracy of the generation of the current: - 5%±20 of [emr] (in the range to 190 ms); - 1,2±6 of [emr] (from 190 ms to 5[s]). Selection of the time behavior: Immediate wear and tear or after the intersection of zero level. Selection of regime on the current: 1/2I * I * 2I * 5I. *) it is i-th the nominal value of current. Selection of the regime of the indication: With 0 - phase (0°) and before reversed phase (180°).
CALIBRATION OF THE TESTERS OF THE GROUNDING Fixed resistances: 0,05; 0,1; 0,22; 0,33; 0,5; 1; 5; 10; 100; 1000 Ohm. Permission: 0,0001 Ohm (to 10 Ohm); 0.001 Ohm (100 Ohm even 1 kilohms). Accuracy: 0,5%± 0,004 Ohm. Automatic correction of the circuit impedance of grounding (option): - The maximum value of the correction: 2 Ohm. - The permission: 0,0001 Ohm. - The accuracy: ±0,018 Ohm. - The measured current: 4 a.
MULTIPROFILE CALIBRATION Groundmeters: - Characteristics you look before p. “the testers of [zazamleniya]”. Gauges of the current of the ground circuits: - The ranges: 100 mA, 10 A, 30 a. - The permission: 1 mA, 10 mA. 10 mA. - The accuracy: 1,5% of ±6 of [emr]. Impedometers of the isolation: - Range 1: from 0 to 5 M, error 0,3%. - Range 2: from 5[MOm] to 2[GOm], error 3%.
Gauges of the leakage current: - The ranges: 2 mA, 4,7 mA. 7,7 mA. - The permission: 1 A. - The accuracy: 1,5% of ±2 of [emr]. Testing units on the stress - The ranges: from 100 [v] to 300 [v] of rms value. - The permission: 0,1 v. - The accuracy: 1,5% of ±9 of [emr]. Impedometers of the load: - To [kharakteristi] you look before corresponding [p].[p]. Instruments for the testing at the point of stress of the flash: - About class 1 - - The range: from 1 to 1,8 sq. - - The permission: 1 v. - - The load resistance: 600 kilohms (2,5 mA; 1,5 kV).
- - The accuracy: 4%±10 of [emr] (on cl. 1 and 2). - About class 2 - - The range: From 2 to 3,6 sq. - - The permission: 1 v. - - The load resistance: 1,2 M (2,5 mA; 3 kV). - - The accuracy: 4%±10 of [emr] (on cl. 1 and 2). Instruments for the testing at the point of current of the flash: - The range: from 1 to 3 mA with the permission of 10[mkA].
- Accuracy 5%
CALIBRATION OF THE GAUGES OF THE LINE VOLTAGE Range: 200 to 260 v. Permission: 0,1 v. Accuracy: 0,8%±6 of [emr].